Spectrofluorimetry |
Spectrofluorimetry - measurement of fluorescence - a form of light emitted by a substance after irradiation at other wavelengths. Origin of PhotoluminescenceAbsorption of visible or UV radiation raises molecule to an excited state
The energy transitions associated with photoluminescence can be represented on Morse
energy surface diagrams.
Energy Changes in FluorescenceThe Morse curves (see above) display energy transitions for an electron, in a covalent bond A-B, excited by hn, then re-emitting the energy as photoluminescence
Stokes' Law Of FluorescenceNet effect of energy transitions depicted and described above is:
As energy is inversely related to wavelength:
|
![]() |
![]() |
An ABSORPTION spectrum is reproducible when scanned on different instruments (Provided
not distorted by inappropriate settings of scan speed, band pass width etc).
But FLUORESCENCE spectra (excitation and emission) are less reproducible; This is because Fluorescence
is not measured relative to a blank.
Hence fluorescence spectra affected by:
Absorption spectra unaffected by the above because absorbance is based on ratio Io/I.
Instrument-related factors affect Io and I in same proportion, so
cancel.
Linear response, Fluorescence vs Mass of Analyte, only at LOW CONCENTRATIONS.
Important factor is the INNER FILTER EFFECT.
![]() |
![]() |
Advantages and Disadvantages of Spectrofluorimetric Assay
Major advantage is high sensitivity.
Capacity to assay at much lower concentrations than by absorption spectrophotometry.
A potential advantage is improved selectivity.
Requirement to set two wavelengths in spectrofluorimetry (excitation and emission) hence
unlikely that an impurity is being co-measured - it would have to absorb and emit at the
same two wavelengths.
Disadvantages are many. A very exacting technique, requiring careful attention to experimental detail, including purity of reagents.
Structural requirements for fluorophore are more difficult to define than for
chromophore.
But a molecule must first absorb hn to undergo the first
step of electron excitation.
Then, because of Stokes' shift to higher wavelength for the emitted light, most
fluorescent species are compounds that absorb UV light. For biological structures,
an aromatic ring is the most common structural requirement. Additional requirement
for fluorescence is that quenching must not occur while the molecule is in the
excited state.In aromatic molecules, electron withdrawing groups tend to produce
quenching. eg -NO2, or -COOH substituents on the aromatic ring tend
to decrease the chances of fluorescence. Conversely, electron rich groups inhibit
quenching. eg -NH2 or -OH are likely to enhance the fluorescence,
e.g., Fluorimetric Assay Of Catecholamines. (Biologically active amines - include
the hormones adrenalin and noradrenalin. Elevated levels in some medical conditions).
![]() |
Product has 4 electron donor groups. Highly fluorescent at 530 nm (360
nm excitation). Basis of sensitive assay for adrenaline.